KROHNE CORIMASS MFC 85 EN User Manual
Page 6
6
3.2
Displaying Concentration and Solute Flow Values
Table 3.2 shows the measured values available for display on the converter with the
concentration systems. All options have a Concentration by Mass display, (in units of °Brix,
°Baumé or just as a percentage), and a solid flow rate display. The General Concentration
option also has a percentage by volume display.
The control of these displays are basically the same for all systems, please refer to the
Standard operating manual. Using NaOH as an example, starting from measuring mode:
Key
Line 1
Line 2
Comments
→↑↑
Fct. 3.(0)
INSTALL
→↑
Fct. 3.(2).0
DISPLAY
→↑
x8
Fct. 3.2.(9)
SOLUTE.FLOW
→
(OFF)
Display disabled
↑
00000.000
(g)/min
Display enabled, use
↑
key
to select mass units
↑
00000.000
(kg)/min
→
00000.000
kg/(min)
Use the
↑
key to select
time units
→
00000(.)000
kg/min
Use the
↑
key move the
decimal point
↑
000000(.)00
kg/min
↵
Fct. 3.2.(9)
SOLUTE.FLOW
↑
Fct. 3.2.(10)
CONC.BY.MASS
→
(OFF)
Display disabled
↑
000000.00
(PERCENT M)
Display enabled
↵
Fct. 3.2.(10)
CONC.BY.MASS
↵
x4
Exit menus and save changes
Table 3.2.1 Measurement Display variations for Concentration systems
Concentration Option
NONE
BRIX
BAUMÉ
GEN. CONC.
NaOH
MASS FLOW
MASS FLOW
MASS FLOW
MASS FLOW
MASS FLOW
MASS TOTAL
MASS TOTAL
MASS TOTAL
MASS TOTAL
MASS TOTAL
DENSITY
DENSITY
DENSITY
DENSITY
DENSITY
TEMPERATURE
TEMPERATURE
TEMPERATURE
TEMPERATURE
TEMPERATURE
VOLUME FLOW
VOLUME FLOW
VOLUME FLOW
VOLUME FLOW
VOLUME FLOW
VOLUME TOTAL
VOLUME TOTAL
VOLUME TOTAL
VOLUME TOTAL
VOLUME TOTAL
BRIX
BAUME
PERCENT M.
PERCENT M.
SUCROSE FLOW
SUCROSE FLOW
PERCENT V.
SOLID FLOW
SOLID FLOW
BOLD type indicates displays that are permanently enabled. All other displays must be
specifically enabled in the corresponding Fct 1.2 or 3.2 menu.