KROHNE IFC 090F Converter EEx-ATEX EN User Manual

Page 6

background image

ALTOFLUX IFC 090 F-EEx and ALTOFLUX IFC 090 F / i-EEx

6

Mains fuses
Power supply

Mains fuse

Type Nominal

voltage

Rating

Specifications

100/115 Vac

200 mA

Switching char.: fast (F) to time-lag (T)

200/230 Vac

125 mA

Breaking capacity:

≥ 1500 A @ 250 V

AC-versions

24 Vac

Switching char.: fast (F) to time-lag (T)

DC-version

24 Vdc

1.25 A

Breaking capacity:

≥ 300 A @ 65 V


The IFC 090-EEx electronics unit is equipped with the following in-/output circuits. Terminals B1,
B

⊥ and B2 can be configured as status or pulse outputs or as control inputs via the software. See

the table below for the electrical data of these in-/output circuits.

Electrical data of in-/output circuits
Terminals Description

Nominal voltage

Maximum current

B1, B

⊥, B2 Pulse, status, control in-/outputs

32 V

150 mA

I+, I

Current output

15 V

22 mA


The difference between 115/230 Vac version and the 100/200 Vac version is the number of the
primary windings of the mains transformer.

The terminals indicated as B1, B

⊥ and B2 can be configured as pulse in-/outputs by jumpers.

The IFC 090-EEx electronics unit is provided with an extension module that is used for data
communication (e.g. SMART, HART or RS485). In that case, the current output signal (I+, I) is
superposed with a sinusoidal voltage signal of 0.5 V. The extension module does not significantly
influence the nominal voltage U

n

, the nominal current I

n

or the power dissipation.


Overheating protection
The mains transformer is protected against overheating by a temperature breaker in series with
the primary winding. The temperature breaker is casted within the compound of the transformer
and has an opening temperature of 125 °C (± 5 K).

IFC 090i-EEx unit with MODIS modules
The IFC 090i-EEx electronics unit is equipped with two MODIS-modules. It is equipped with one of
the following power supplies.

Electrical data of IFC 090i-EEx electronics unit
Power supply

Terminal

Function

Electrical data

AC-version

L
N
PE

Live
Neutral
Protective ground

U

n

= 100…230 Vac –15%/+10%

P

n

= approx. 15 VA


Primary fuse: Rated current: 1.6 A

Design conform IEC 127-2

Breaking capacity: 1500 A

AC/DC-version 1L

0L
FE

Live
Neutral
Functional ground

U

n

= 24 V ac / dc

AC:

-15%/+10% or 20.4…26.4 V ac

DC:

-25% / +30% or 18…32 V dc

I

n

=

approx. 490…630 mA for AC

approx. 310…550 mA for DC

P

n

= approx. 10 W


Primary fuse: Rated current: 1.25 A

Design conform IEC 127-2

Breaking capacity: 1500 A


NOTE:

The mains fuses for both electronics units are listed in Sect. 6 of this manual.