Description of hart interface – KROHNE SMARTSENS ORP 8150 EN User Manual
Page 34
7
DESCRIPTION OF HART INTERFACE
34
SMARTSENS ORP 8150
www.krohne.com
11/2013 - 4002546702 - MA SMARTSENS ORP 8150 R02 en
4 Setup
1 Process input
Temperature
Set temperature unit °C / °F
Set temperature offset
Display date of offset calibration
Calibration
Start calibration procedure. For more information
refer to
Calibration on page 17.
Maintenance interval
Set maintenance interval in days (000...999 days);
Default setting: 000
Reset maintenance
interval
Yes / No; Default setting: No
2 I/O
Meas. value at 4 mA
Set measuring value at 4 mA; Default setting:
-1500 mV
Meas. value at 20 mA
Set measuring value at 20 mA; Default setting:
+1500 mV
Time constant
Set time constant (1…60 seconds); Default
setting: 1 second
3 I/O HART
TAG
Set TAG for measuring loop
Set long TAG
Display of previous long TAG
Message
Display messages (32 packed ASCII)
Polling address
Display polling address
Loop current mode
Select between enable or disable to activate or
deactivate the loop current mode
Device variables
PV
Display PV - ORP with
min. and max. limits
SV
Display SV - electrode
voltage in mV
TV
Display TV -
temperature in °C / °F
with min. and max.
limits
.book Page 34 Monday, November 18, 2013 10:29 AM