beautypg.com

Technical data, 2 technical data – KROHNE SMARTSENS ORP 8150 EN User Manual

Page 25

background image

TECHNICAL DATA

6

25

SMARTSENS ORP 8150

www.krohne.com

11/2013 - 4002546702 - MA SMARTSENS ORP 8150 R02 en

6.2 Technical data

INFORMATION!

The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.

Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).

Measuring system

Measuring principle

Potentiometric

Measuring range

-1500...+1500 mV

Design

Shaft diameter

12 mm / 0.47"

Insertion length

120 mm / 4.72"

Temperature sensor

Pt1000

Connector

VarioPin 2.0 (VP 2.0)

Operating conditions

Temperature range

0...+130°C / +32...+266°F

Process pressure

12 bar / 174 psi

Measuring accuracy

0.5%

Conductivity

Min. 150 µS/cm

Installation conditions

Ingress protection

IP 68

Weight

Approx. 80 g / 0.18 lb

Process connection

PG 13.5

Materials

Sensor shaft

Glass

Measuring electrode

Platinum

Inner buffer

pH 7.0

Reference

Duralid gel

Diaphragm

2 x open diaphragm

Gasket

EPDM (FDA)

Sensor head

Nickel plated brass body with VP 2.0 connector

PEEK body with VP 2.0 connector (in preparation)

.book Page 25 Monday, November 18, 2013 10:29 AM