DESCRIPTION OF HART INTERFACE
7
33
SMARTSENS ORP 8150
www.krohne.com
11/2013 - 4002546702 - MA SMARTSENS ORP 8150 R02 en
7.8 Basic-DD menu tree (details for settings)
1 Measuring values
1 ORP value
Display of the measured ORP value on process control system, HART
®
handheld or display.
2 Temperature
Display of the measured temperature value in °C / °F on process control
system, HART
®
handheld or display.
3 Electrode voltage
Display of the measured electrode voltage in mV on process control system,
HART
®
handheld or display.
4 Loop current
Display of the measured loop current in mA on process control system,
HART
®
handheld or display.
5 Error
Display of status messages and diagnostic information. For further
information refer to
Status messages and diagnostic information on page 20.
6 Device state
Display of status information icon for sensor according to NAMUR NE 107.
2 Quick setup
1 TAG
Set the TAG number for the measuring loop.
2 Manual hold
Activate or deactivate the manual hold function. Select between yes or no.
3 I/O
Meas. value at 4 mA
Set measuring value at 4 mA;
Default setting: -1500 mV
Meas. value at 20 mA
Set measuring value at 20 mA;
Default setting: +1500 mV
Time constant
Set time constant (1…60 seconds);
Default setting: 1 second
4 Calibration
Start calibration procedure. For more information refer to
Calibration on
page 17.
5 Input user password
Set password
3 Logbooks
1 Calibration logbook
Logbook for the last 20 calibration cycles
2 Error logbook
Display errors