INSTALLATION

3

15

SMARTSENS ORP 8150

www.krohne.com

11/2013 - 4002546702 - MA SMARTSENS ORP 8150 R02 en

Figure 3-4: SJB 200 W and SD 200 W/R (left side). SJB 200 W for HART

®

communication and with resistance R (right

side).

SJB 200 W and SD 200 W/R

SJB 200 W for HART

®

communication and

with resistance R

A

Sensor +

B

Sensor +

C

Sensor -

C

Sensor -

D

Display + (SD 200 W/R)

G

PLC - (process control system)

E

Display - (SD 200 W/R)

H

PLC + (process control system)

F

PLC - (process control system)

I

PLC + (process control system)

Figure 3-5: SJB 200 W for HART

®

communication without resistance R (left side). SJB 200 W (right side).

SJB 200 W for HART

®

communication and without

resistance R

SJB 200 W

A

Sensor +

A

Sensor +

C

Sensor -

C

Sensor -

G

PLC - (process control system)

G

PLC - (process control system)

I

PLC + (process control system)

I

PLC + (process control system)

.book Page 15 Monday, November 18, 2013 10:29 AM