Operation, 3 status messages and diagnostic information – KROHNE SMARTSENS ORP 8150 EN User Manual

Page 20

background image

4

OPERATION

20

SMARTSENS ORP 8150

www.krohne.com

11/2013 - 4002546702 - MA SMARTSENS ORP 8150 R02 en

4.3 Status messages and diagnostic information

Measurements out of specification

Maintenance

Message

Message

Message

Message

Description

Description

Description

Description

Action

Action

Action

Action

S01

ORP value > ORP max
ORP value out of range

Consider the measuring range limits, otherwise
select a suitable sensor for the process conditions of
the application.

S02

ORP value < ORP min
ORP value out of range

S03

Temp. value > Temp max
Temperature value out of range

S04

Temp. value < Temp min
Temperature value out of range

Message

Message

Message

Message

Description

Description

Description

Description

Action

Action

Action

Action

M01

Offset < -99 mV

Recalibrate the sensor. For more information refer
to

Calibration with PACTwareTM on page 17

M02

Offset> +99 mV

M03

Maintenance interval expired

.book Page 20 Monday, November 18, 2013 10:29 AM