Peak values
When using sensors working acc. to the non-contact
measuring principle, the measurement can be influen-
ced by the respective process conditions. In this menu
item, the reliability of the level echo is displayed in dB.
The reliability is signal strength minus noise. The higher
the value, the more reliable the measurement.
Indication "OK" or flashing error message, e.g. "E013".
In addition, the error appears in clear text in the
measured value display.
Meas. reliability
Sensor status
The signal curves enable a first rough evaluation of the
measurement. The following curves are available:
l
Echo curve
l
False echo curve
l
Trend curve.
The echo curve shows the echoes with signal strength in
db over the distance.
The false echo curve shows the saved false echoes
(see menu "Service") of the empty vessel with signal
strength in "dB" over the measuring range.
Choose curve
With a comparison of echo and false echo curve a more
detailed statement of the reliability can be made. The
selected curve is updated permanently. With the [OK]
key, a submenu with zoom functions is opened.
Available with the echo and false echo curve:
Meas. reliability
Sensor status
Choose curve
Curve presentation
16