INSTALLATION

2

7

OPTIFLEX 1100 C

www.krohne.com

01/2012 - 4001226201 - QS OPTIFLEX 1100 R01 en

2.2.2 How to assemble the segmented coaxial probe

• Put the signal cable into the first segment of tube. Make sure that the signal cable goes

through the center hole in the spacer.

• Attach the first segment of tube to the thread at the top of the signal cable. Tighten with a

17 mm open-end wrench.

• Attach the lock screw to the top of the tube. Tighten with a 3 mm Allen wrench.
• Do these steps again for each segment of tube, but do not attach the lock screw to the last

segment of tube.

Attach the tube segments

Figure 2-5: How to assemble the segmented coaxial probe: part 1

1 Top segment of tube
2 Bottom segment of tube
3 1 HC M6×5 lock screw
4 Bonded plastic spacer

CAUTION!

After you tighten the lock screw on each segment of tube, it will not be possible to
disassemble the coaxial probe.

If it is necessary to adjust the probe length, do not attach the lock screw to the last segment
of tube!