1.3

Key for Pressure Equipment Directive

PED

/ / /

1 2 3

4

5


1

P

ressure Equipment Directive


2

Fluid

G Gases, liquefied gases, gases dissolved under pressure, vapors and those liquids

whose vapor pressure lies more than 0.5 bars over the normal atmospheric pressure

(1013 mbars) at the maximum permissible temperature

L

Liquids whose vapor pressure lies a maximum of 0.5 bars above the atmospheric

pressure at the maximum permissible temperature

3

Fluid group

1

Group 1: Explosion-hazardous, highly flammable, readily flammable, flammable

(when the maximum permissible temperature lies above the flash point),

highly toxic, toxic, fire stimulating

2

Group 2: All the fluids not specified in Group 1

4

Category

3.3 In accordance with Article 3.3 of Directive 97/23/EC

I

Category I to 97/23/EC

II Category II to 97/23/EC

III Category III to 97/23/EC

5

Conformity evaluation process

SEP Solid engineering practice

A Module A internal process inspection

A1 Module A1 internal process inspection with supervision of the acceptance

H Module H Comprehensive quality assurance


The PED key marking is contained on the rating plate of the instrument.





1.4 Functional

principle

A half-round plate, or flap, is fastened across the
direction of flow to a springmounted rotating spindle
in a ring that is inserted between flanges in a pipeline.
As the flow rate increases, the flap rotates counter
to the restoring force F of the spring in the direction
of flow.
The ensuing angle of rotation, depending on the
volume rate of flow, is transmitted via a magnetic
coupling to the indicator part.