ELECTRICAL CONNECTIONS

4

27

OPTISONIC 7300 / 8300

www.krohne.com

05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02

4.3 "Ex i" I/O connections

The following intrinsically safe I/O connections are available:

The I/O circuits titled “Ex i IO” , "Ex i Option" and “Ex i Option 2” are always provided with type of
protection Intrinsic Safety (Ex ia). The I/O-circuits “Fieldbus I/O Profibus-PA” as well as “Fieldbus
I/O Foundation Fieldbus” can be provided with type of protection Intrinsic Safety.

Up to a maximum of 4 intrinsically safe (Ex ia) in-/outputs are possible. All intrinsically safe
circuits are galvanically isolated from earth and each other. To avoid summation of voltages and
current, the wiring of these “Ex ia”-circuits must be sufficiently separated, e.g. in accordance
with the requirements of standard EN 60079-14, clause 12.2.
The “Ex ia” in-/outputs may only be connected to other “Ex ia” or “Ex ib” approved devices (e.g.
intrinsically safe isolation amplifiers), even if such devices are installed in a non-hazardous
location!
Connection to non-“Ex i” devices cancels the “Ex ia” properties of the flowmeter.

Terminals L and N (or L+ and L-) for connection of the mains supply are not available with type of
protection “intrinsic safety”. To achieve the necessary separation distances according to EN
60079-11 between the non-”Ex i” and “Ex i” circuits, the mains terminals are provided with a
semi-circular protection cover with a “snap-in” lock. This cover MUST be closed before
establishing the power supply to the converter.

I/O PCB

I/O functions

Ex i IO

Current Output, passive + HART

communication

Pulse / Status Output, passive

Ex ia IIC or Ex ib IIC

U

i

= 30 V, I

i

= 100 mA, P

i

= 1.0 W

C

i

= 10 nF, L

i

= negligibly low

Current Output, active + HART

communication

Ex ia IIC or Ex ib IIC

U

o

= 21 V, I

o

= 90 mA, P

o

= 0.5 W

Lineair characteristics

C

o

= 90 nF, L

o

= 2.0 mH

C

o

= 110 nF, L

o

= 0.5 mH

Ex i Option

Current Output, passive

Pulse / Status Output / Control

Input, passive

Ex ia IIC or Ex ib IIC

U

i

= 30 V, I

i

= 100 mA, P

i

= 1.0 W

C

i

= 10 nF, L

i

= negligibly low

Current Output, active

Ex ia IIC or Ex ib IIC

U

o

= 21 V, I

o

= 90 mA, P

o

= 0.5 W

Lineair characteristics

C

o

= 90 nF, L

o

= 2.0 mH

C

o

= 110 nF, L

o

= 0.5 mH

Ex i Option 2

Current Input, passive

Pulse / Status Output / Control

Input, passive

Ex ia IIC or Ex ib IIC

U

i

= 30 V, I

i

= 100 mA, P

i

= 1.0 W

C

i

= 10 nF, L

i

= negligibly low

Current Input, active

Ex ia IIC or Ex ib IIC

U

o

= 24.1 V, I

o

= 99 mA, P

o

= 0.6 W

Lineair characteristics

C

o

= 75 nF, L

o

= 0.5 mH

Fieldbus I/O

Profibus-PA

Foundation Fieldbus, passive

Ex ia IIC or Ex ib IIC / IIB

U

i

= 24 V, I

i

= 380 mA, P

i

= 5.32 W

C

i

= 5 nF, L

i

= 10 µH

Suitable for connection to an

intrinsically safe fieldbus in

accordance with the FISCO-

model