beautypg.com

Introduction, 9 technical data 1.10 marking codes – KROHNE OPTISONIC 7300 Ex EN User Manual

Page 10

background image

1

INTRODUCTION

10

OPTISONIC 7300 / 8300

www.krohne.com

05/2012 - 7312262200 - AD EX OPTISONIC 7300/8300 R02

1.9 Technical data

1.10 Marking codes

Due to the various configuration options, the OPTISONIC 7300 gas flowmeter series can have
several (different) explosion safety measures implemented.

Process medium

Flammable and non-flammable gases

Protection category acc. to IEC 529 /

EN 60529

IP 65 or better

Ambient temperature range

Ambient temperature range

Ambient temperature range

Ambient temperature range

Titanium transducers

OPTISONIC 7300 C/...-Ex: -40...+65°C

1

GFC 300 F/...-Ex: -40...+65°C

1

OPTISONIC 7000...-Ex (incl. XT-version) : -40...+70°C

Stainless steel HT / steam transducer

OPTISONIC 8000-Ex: -40...+60°C

Synthetic transducers

2

(only available in OPTISONIC 7000-Ex)

Type G5.nn: -50...+70°C

Type G6.nn, Temp class T6: -50...+70°C

Type G6.nn, Temp class T5: -50...+85°C

Type G6.nn, Temp class T4: -50...+100°C

Process medium temperature range

Process medium temperature range

Process medium temperature range

Process medium temperature range

Titanium transducers

OPTISONIC 7300 C/...-Ex: -50...+180°C

OPTISONIC 7000-Ex: -50...+180°C

OPTISONIC 7000 XT-Ex: -50...+220°C

Stainless steel HT / steam transducer

OPTISONIC 8000-Ex: -200...+440°C

Synthetic transducers

2

(only available in OPTISONIC 7000-Ex)

Type G5.nn, Temp. class T6: -50...+70°C

Type G6.nn, Temp class T6: -50...+70°C

Type G6.nn, Temp class T5: -50...+85°C

Type G6.nn, Temp class T4: -50...+100°C

INFORMATION!

1

Maximum ambient temperature applies to aluminum converter housing. Stainless steel

version is limited to +55

°

C.

2

The ATEX-certificate KEMA 07 ATEX 0181 X (Issue no. 3) describes the maximum ambient

temperatures of transducers type G5.nn and G6.nn, which should be read as process
temperature (this is the temperature of the medium that surrounds the transducers). The
ambient temperature is the (air) temperature at the location where the meter is installed.