beautypg.com

Electrical installation – KROHNE ALTOSONIC V Installation EN User Manual

Page 22

background image


22 www.krohne.com

04/2012

ALTOSONIC V

ELECTRICAL INSTALLATION

3

Connection parameters per type for UFP card MP103

Analog inputs

AD812 card
PCLD 780 wiring board

Pmax (limit for used 250 ohm resistors) = 0.2 W
Maximum Overvoltage continuous =+/- 7V
Usually connected through barriers

Analog
outputs

AD812 card
PCLD 780 wiring board

Maximum Output drive =5mA , 10V max
Normally connected through
barrier converters(0..10V :: 4-20mA)