RD-8X-N4X Operators Manual

This is pre-programmed to the value of 50 and should not be altered unless directed to do so by Detcon
Factory Personnel.

Use the “u” (Up Arrow) and “v” (Down Arrow) keys to move from one parameter to another. When the
Blinking Cursor is positioned at the parameter that needs to be modified, the “ENTER” key is used to open
that parameter to change. Use the “u” (Up Arrow) and the “v” (Down Arrow) keys to change the
parameter to the desired value, and press the “ENTER” key to accept that value.

NOTE: The “TYPE” parameter is an alphanumeric display of 10 characters that can be changed
by use of the “Arrow” keys (“u”,“v”). When the appropriate alphanumeric character is found the
“ENTER” key is used to accept that character, and move to the next character. Once all ten
characters have been selected (including spaces or blank characters) the display will return to the
Blinking Cursor.


When the Blinking Cursor is at the end of a set of parameters for one channel, the “Down Arrow” key will
move the Blinking Cursor to the next channel’s set of parameters.

Press the “PROG” key to exit the “SET CHANNEL DATA” menu and return to the Main Manu, press the
“PROG” key again to exit the Main Menu or press the “v” (Down Arrow) key to move to the next menu
item, or the “u” (Up Arrow) to move to the previous menu item.

4.2.3 Calibration Mode

“CALIBRATION MODE” disables all alarms until the calibration mode is exited.

To enter the “CALIBRATION MODE” Menu item, press the “PROG” key to enter the Main Menu. From
the Main Menu, press the “v” (Down Arrow) key until the display shows “CALIBRATION MODE”.
Press the ENTER” key to enter calibration mode.

The relays and alarm LED’s will remain disabled until the “ENTER” key is pressed again,
“CALIBRATION MODE” is exited, and the unit returns to normal operation.

NOTE: The RD-8X does not poll sensors or control systems while in Menu Mode, so there are no
updates to gas readings or alarms.


Press the “PROG” key to exit the Main Manu, or press the “u” (Up Arrow) key to move to the previous
menu item.

4.3 Remote Display Setup


In order for the model RD-8X to communicate with other Detcon devices, certain channel parameters must
be configured. The following tables below give the needed channel setup parameters for the model RD-
8X.

4.3.1 Model RD-8X to Model 1600A or RD-8X

Table 1

RD-8X: Setup Channel Data

Model 1600A / RD-8X

Channel

Slave ID

Register to Read

Reading Register

Channel

1 **

16

0

1

2 **

16

1

2

Model RD-8X-N4X Operator Manual

Rev. 1.0

Page 7 of 14