activated. Typically, air flow should range between 500 and 1000 ml/min. (milliliters per minute) as seen
on the flow meter located on the left of the enclosure. It is recommended that the flow fault circuit be set to
activate at a flow of approximately 300 ml/m.

To adjust the flow fault first set the air flow at the point where you want the fault condition to activate.

Next turn the flow fault adjust potentiometer (PT1 as seen in the illustration on the previous page) clock-
wise until the flow fault LED illuminates and its relays fire. Make adjustments as necessary by increasing
and decreasing airflow until the flow fault function occurs at the desired set point. Once the flow fault cir-
cuit has been set, return the air flow to between 500 and 1000 ml/m.

NOTE: Because sample and vent tubing can become congested over a period of time, the flow fault func-
tion should be checked periodically and adjusted if necessary.

1.4 CALIBRATION

Model 972-071170-0LC sensor assemblies are calibrated prior to shipment and should therefore require
minimal adjustment at time of commissioning. However, it is recommended that a complete calibration be
performed on a periodic basis to assure optimum system performance. Refer to the sensor instruction man-
ual for gas type, range and sensor specific calibration instructions.

Note that the sensor instruction manual requires the removal of rain/splash guards for the application

of gas. This requirement will be superceded by the instructions below. The calibration adapter attached to
the sensor head should not be removed during calibration.

Required material:
1. Span gas cylinder containing gas as per instructions in the sensor instruction manual.
2. 1/4" Tee tubing fitting as shown in the illustratin below.

Detcon Model 972-071170-0LC Gas Detection System PG.5

6" of
vent
tube

Calibration

Port

Fixed flow

regulator

Span gas

cylinder

Teflon
tubing

1/4"
Tee

400 ml/m

500
ml/m

0

100

200

300

400

500

detcon inc.
HOUSTON, TEXAS

SAMPLE DRAW

GAS DETECTION SYSTEM

SET

FLOW

CAL

IN

CL

OSE

O P E N

SAMPLE

VALVE

CL

OSE

O P E N

VENT

100 ml/m

NOTE: The 1/4'' Tee and vent tube
is not necessary when using a
demand flow regulator. Set flow via
rotameter to 500 cc/min.