Detcon 1000_H2S User Manual
Page 25

Model 1000 H2S Echem
Model 1000 H2S Echem Instruction Manual
Rev. 2.8
Page 21 of 30
Sensor life is an estimated remaining use of the sensor head, between 0% and 100%
Example: 85=85% sensor life
Register #
High Byte
Low Byte
40001
Detectable
Range
i.e. 100 for 0-100 ppm, 50 for 0-50% LEL, etc
Register #
High Byte
Low Byte
40002
Current Gas Reading
The current gas reading as a whole number: If the reading is displayed as 23.5 on the display, this register
would contain the number 235.
Register #
High Byte
Low Byte
40003
Alarm 1 Set point
This is the trip point for the first alarm.
Register #
High Byte
Low Byte
40004
Alarm 2 Set point
This is the trip point for the second alarm.
Register #
High Byte
Low Byte
40005
Status Bits Status Bits
High Byte
Bit 7
Not used, always 0
Bit 6
Not used, always 0
Bit 5
Not used, always 0
Bit 4
Not used, always 0
Bit 3
1-Unit is in calibration 0-Normal operation
Bit 2
1-Alarm 2 is ascending 0-Alarm 2 is descending
Bit 1
1-Alarm 2 is normally energized 0-Alarm 2 is normally de-energized
Bit 0
1-Alarm 2 is latching 0-Alarm 2 is non-latching
Low Byte
Bit 7
1-Alarm 2 Relay is energized 0-Alarm 2 Relay is not energized
Bit 6
1-Alarm 1 is ascending 0-Alarm 1 is descending
Bit 5
1-Alarm 1 is normally energized 0-Alarm 1 is normally de-energized
Bit 4
1-Alarm 1 is latching 0-Alarm 1 is non-latching
Bit 3
1-Alarm 1 Relay is energized 0-Alarm 1 Relay is not energized
Bit 2
1-Fault is normally energized 0-Fault is normally de-energized
Bit 1
1-Fault is latching 0-Fault is non-latching
Bit 0
1-Fault Relay is energized 0-Fault Relay is not energized
The following is a typical Master Query for device # 8:
Field Name
HEX
DEC
RTU
Slave Address
08
8
0000 1000
Function
03
3
0000 0011
Start Address Hi
00
0
0000 0000