beautypg.com

Calibration and maintenance – Detcon 1000_H2S User Manual

Page 20

background image

Model 1000 H2S Echem

Model 1000 H2S Echem Instruction Manual

Rev. 2.8

Page 16 of 30

5.2.4

Software Flow Chart

Figure 14 DM-624 Software Flow Chart

6. Calibration and Maintenance

The Model 1000 Series H

2

S Process Analyzer is calibrated prior to shipment. Only minimal adjustment

should be required at time of commissioning. However, it is highly recommended that a zero and span
calibration be performed twice in the first week of operation to assure optimum system performance. After
that, a recalibration interval of every month is recommended.

Maintenance

When the Acetic Acid Bubbler option is used, the 5% acetic acid solution (or white vinegar) should be kept
near the Max fill mark on the bottle. The level should be checked monthly and the entire solution contents
should be changed every 6 months.