beautypg.com

Figure 5 calibration setup, Figure5 calibration setup, Houston,texas – Detcon 971-071170-0DL User Manual

Page 9

background image

Model 971-071170-0DL

Note:

It is critical that air dilution be set in coordination with the entry value for AutoSpan.

The air dilution used during span calibration must be maintained as a constant or calibration
will drift. If air dilution ratio is changed, span calibration must be repeated.

detcon, inc.

HOUSTON,TEXAS

SAMPLE DRAW

GAS DETECTION SYSTEM

CAL PORT

SET
AIR

FLOW

SET

SAMPLE

FLOW

SAMPLE

PORT

0

50

100

150

200

250

0

0.4

0.8

1.2

1.6

2.0

1.0

0.2

0.6

1.4

1.8

Fixed Flow
Regulator

6" of vent tube

CALIBRATION

PORT

Teflon tubing

Span Gas

Cylinder

1000

ml/min

500 ml/min

Figure5 Calibration setup

Figure5


Required material:

1. Span gas cylinder containing gas as per instructions in the sensor instruction manual.
2. 1/4" Tee tubing fitting as shown in

.

To Calibrate: (refer to Figure5)

1) Turn the 3-way valve so that the arrow is pointing down. Sample is now being drawn from the CAL port.
2) a) Using the 1/4" Tee assembly, connect the span gas cylinder to the calibration port. Note that the

cylinder flow rate should be 1000 ml/m while the sample draw flow rate should be adjusted from 100- 500
ml/m. This will assure that the sample is not diluted. Be sure to add the 6” vent tubing as shown to
prevent dilution.
b) Make sure air dilution flow is set to target level according to customer’s requirement.

3) Follow the calibration instructions given in the sensor instruction manual.
4) Close cylinder valve to allow clearing/zero stability. Note that zero and span functions may interact

slightly with one another. Zero and span tests may be repeated 2 or 3 times as necessary.

5) When calibration is complete turn the 3-way valve so that the arrow is pointing up.

971-071170-0DL Instruction Manual

Rev. 0.0

Page 5 of 8