Model 971-071170-0DL

2.2 Sampling connections

1. Install a length of tubing from the desired sample point to the sample port as shown in.

Sample draw tubing should be either Teflon or stainless steel 1/4" OD. Sample flow rates are
set between 500 and 1000 ml/m.

Note:

When detecting for NH3, CL2, CLO2, or HCL care must be taken to use a minimum amount of

stainless steel tubing and connections because stainless steel readily absorbs these gasses and will
cause significant error in the detection process. Teflon tubing is recommended for these applications.

detcon, inc.

HOUSTON,TEXAS

SAMPLE DRAW

GAS DETECTION SYSTEM

CAL PORT

SET

AIR

FLOW

SET

SAMPLE

FLOW

SAMPLE

PORT

Sample

Port

0

50

100

150

200

250

0

0.4

0.8

1.2

1.6

2.0

1.0

0.2

0.6

1.4

1.8

Figure 2 Dilution Flowmeter Bracket

2. Install a length of tubing from the vent port and an area deemed safe for vent as shown below.

VOC Sensor

MODEL

PI-700

PGM1

ZERO

PGM2

SPAN

Vent Port

VOC Sensor

MODEL

PI-700

PGM1

ZERO

PGM2

SPAN

Figure 3 Sensor assembly vent port

971-071170-0DL Instruction Manual

Rev. 0.0

Page 2 of 8