9.0 TROUBLE SHOOTING
Memory or Error Reports
1. Reinitialize Sensor Unplug transmitter and replug transmitter then swipe magnet over PGM 1 in the
first 3 seconds. This will clear the processor and recover from error state. Remember to put in all
customer settings for range, alarm and cal gas level after re-initialization.
Non-readable Display
1. If display has blue background when hot, install sunshade to reduce temperature.
2. If poor contrast, adjust contrast pot accordingly.
Nothing Displayed – Transmitter not responding
1. Verify condulet has no accumulated water or abnormal corrosion.
2. Verify required DC power is applied to correct terminals.
3. Swap with a known good transmitter to determine if transmitter is faulty.
Bad 4-20 mA Output or RS485 Output
1. Check that wiring is connected to correct terminal outputs.
2. Swap with a known good transmitter to determine if transmitter is faulty.