Model TP-700

TP-700 Instruction Manual

Rev. 3.5

Page 8 of 38

When mounting on a pole, secure the Junction Box to a suitable mounting plate and attach the mounting plate
to the pole using U-Bolts. (Pole-Mounting brackets for Detcon J-box accessories are available separately.)

2.5 Electrical Installation

The Sensor Assembly should be installed in accordance with local electrical codes. The sensor assemblies are
CSA/NRTL approved (US and Canada) for Class I, Division 1, Groups B, C, & D area classifications, and are
ATEX Approved for Class I, Zone 1, Group IIB+H

2

area classifications.

Proper electrical installation of the gas sensor is critical for conformance to Electrical Codes and to avoid
damage due to water leakage. Refer to Figure 7 and Figure 8 for proper electrical installation.

NOTE: If a conduit run exits the secondary port, repeat the installation technique shown in
Figure 7.

In Figure 7, the drain allows H

2

O condensation inside the conduit run to safely drain away from the sensor

assembly. The electrical seal fitting is required to meet the National Electrical Code per NEC Article 500-3d
(or Canadian Electrical Code Handbook Part 1 Section 18-154). Requirements for locations of electrical seals
are covered under NEC Article 501-5. Electrical seals also act as a secondary seal to prevent water from
entering the wiring terminal enclosure. However, they are not designed to provide an absolute watertight seal,
especially when used in the vertical orientation.

Plug any unused

ports

Explosion Proof

Housing

(J-Box)

Drain

TP-700

Sensor

Assembly

Conduit

"T"

EYS Seal Fitting

MODEL

TP-700

detcon inc.

H

2

S Sensor

(+

)

m

A

(-

)

A

(+

)

B

(-

)

Wiring to

Sensor Assembly

W

h

t

B

lu

R

e

d

G

rn

B

lk

Explosion

Proof

Junction Box

(+)

mA

(-)

N/U

A(+)

B(-)

Customer

Supplied Wiring

Transient Protection Module

(TPM) P/N 500-003087-100

Mount TPM in Explosion

Proof Enclosure to ground

unit properly. Mount to

bottom of enclosure using

6-32 screws.

6-Pin Pheonix Plug

P/N 306-175705-100

Figure 7 Typical Installation