DeFelsko PosiTector 200 v.1.2 User Manual
Page 17
![background image](/manuals/366641/17/background.png)
Page 17
T
T
roubleshooting
roubleshooting
Gage does not turn on
Make sure the + and - battery terminals are positioned properly and that fresh
Alkaline batteries are being used.
Gage powers up but fails to stay on
Replace batteries with fresh Alkaline batteries. If problem persists return Gage for
service.
Gage measurements are much lower than expected
Gage may be measuring surface roughness. Raise the value of Gate A. See pg.6
Changing the Batteries
Changing the Batteries
As the batteries become weak the
symbol will appear and begin flashing
during measuring. The batteries should be changed at the earliest opportunity. The
symbol will remain on without flashing when the batteries are very weak.
Gage measurements are much higher than expected
Gage may be measuring both the coating and substrate. Lower the value of Gate
B. See pg.6
Gage does not yield accurate or consistent results
See the Set Gates (pg.6) and Adjustment (pg.7) sections to ensure the gage has
been optimized for your application. Check the Gage on traceable standards.
Make sure to hold the probe in the air and
ensure the probe is free of couplant. If problem
persists, note the error message and contact
our technical support department.
Gage displays an error message while attempting probe ZERO