DeFelsko PosiTector 200 v.1.2 User Manual
Page 10

Normally the Gage powers-down after approximately 60 seconds of no activity.
When SLOW OFF is selected the Gage powers-down after approximately 30 minutes
of no activity. Additionally, an OFF button is placed in the menu. The Gage can
always be powered-down at any time by holding both buttons for 5 seconds.
Page 10
RESET restores factory settings and returns the Gage to a known, out-of-the-box
condition. It is handy when you want to “start all over”. When RESET is selected, the
following screen appears:
The following changes are made to the Gage:
- all stored measurements are erased.
- the calibration adjustment is returned to the Gage’s original factory setting for
polymer.
- Gates are returned to Gate A=25um (1 mil) and Gate B=1000um (40 mil)
- menu settings are changed to the following:
MEMORY = OFF
SLOW OFF = OFF
FLIP LCD = Normal
A “hard” RESET can be performed when the Gage is powered down. In the unlikely
event the Gage fails to power-up or operate properly, hold the button until the
RESET symbol appears.
This performs the same function as a menu RESET with the addition of UNITS = inch
and LANGUAGE = English.
The probe must always be zeroed after a RESET using the ZERO menu option
(pg.6).
Setup Menu Option
Setup Menu Option
This menu option displays the Gage model number, serial number, revision num-
ber and the amount of remaining memory for storage of readings.
When SETUP is selected the following options are presented: