Normally the Gage powers-down after approximately 60 seconds of no activity.
When SLOW OFF is selected the Gage powers-down after approximately 30 minutes
of no activity. Additionally, an OFF button is placed in the menu. The Gage can
always be powered-down at any time by holding both buttons for 5 seconds.

Page 10

RESET restores factory settings and returns the Gage to a known, out-of-the-box
condition. It is handy when you want to “start all over”. When RESET is selected, the
following screen appears:

The following changes are made to the Gage:
- all stored measurements are erased.
- the calibration adjustment is returned to the Gage’s original factory setting for

polymer.

- Gates are returned to Gate A=25um (1 mil) and Gate B=1000um (40 mil)
- menu settings are changed to the following:

MEMORY = OFF
SLOW OFF = OFF
FLIP LCD = Normal

A “hard” RESET can be performed when the Gage is powered down. In the unlikely
event the Gage fails to power-up or operate properly, hold the button until the
RESET symbol appears.

This performs the same function as a menu RESET with the addition of UNITS = inch
and LANGUAGE = English.

The probe must always be zeroed after a RESET using the ZERO menu option
(pg.6).

Setup Menu Option

Setup Menu Option

This menu option displays the Gage model number, serial number, revision num-
ber and the amount of remaining memory for storage of readings.

When SETUP is selected the following options are presented: