COSA Xentaur Xentaur XDTL-HT User Manual
Page 48

48
Register
Number
Address
Access
Level
Data
Type
Description
Values
40405
404 (194h)
Admin
16-bit
Analog output 1 variable
0=Off
64=Dewpoint
65=Sample temperature
66=Sample pressure
67=Enclosure pressure
68=User input 1
69=User input 2
70=User input 3
71=User input 4
120=PID 1
121=PID 2
122=PID 3
123=PID 4
32768=Remote controlled
40406
405 (195h)
Admin
Float
Analog output 1 low limit
Valid values are -100 - 99999.9
40408
407 (197h)
Admin
Float
Analog output 1 high limit
Valid values are -100 - 99999.9
40410
409 (199h)
Admin
16-bit
Analog output 1 remote controlled
value
Valid values are 0 - 65535
40411
410 (19Ah)
Read-Only
16-bit
Analog output 2 type
0=0-5V
1=0-10V
2=±5V
3=±10V
5=4-20mA
6=0-20mA
7=0-24mA
40412
411 (19Bh)
Admin
16-bit
Analog output 2 low limit DAC value
Valid values are 0 - 65535
40413
412 (19Ch)
Admin
16-bit
Analog output 2 high limit DAC value
Valid values are 0 - 65535
40414
413 (19Dh)
Admin
16-bit
Analog output 2 during calibration
operation
0=Set to low state
1=Hold output
2=Track calibration measurement