Setup menu – COSA Xentaur Xentaur XDTL-HT User Manual
Page 27

Analog Outputs
27
Setup Menu
Analog Output
Channel 1
Variable
Lower Range
Upper Range
Type
Calibrate
Test
Lower Limit
Upper Limit
Channel 2
Variable
Lower Range
Upper Range
Type
Calibrate
Test
Lower Limit
Upper Limit
During Cal
During Cal
Channel 2 menu would not appear if the instrument is configured with only one analog output channel.
Select measurement on analog output channel X Go to Analog Output
>
Channel X
>
Variable
Set measurement range on analog output
Go to Analog Output
>
Channel X
>
Lower Range
Set the measurement bottom range
Go to Analog Output
>
Channel X
>
Upper Range
Set the measurement top range
Change the type of analog output on channel X
Go to Analog Output
>
Channel X
>
Type
Test analog output by forcing it to a fixed value
Go to Analog Output
>
Channel X
>
Calibrate
>
Test
Calibrate the digital-to-analog converter
Go to Analog Output
>
Channel X
>
Calibrate
>
Test
Set value to 0%
Go to Analog Output
>
Channel X
>
Calibrate
>
Lower Limit
Set lower limit to a value so that the output at
process monitor is exactly at its bottom range
Go to Analog Output
>
Channel X
>
Calibrate
>
Test
Set value to 100%
Go to Analog Output
>
Channel X
>
Calibrate
>
Upper Limit
Set upper limit to a value so that the output at
process monitor is exactly at its top range