COSA Xentaur Xentaur XDTL-HT User Manual
Page 44

44
Register
Number
Address
Access
Level
Data
Type
Description
Values
40122
121 (79h)
Read-Only
Float
Sample pressure high limit
Valid values are 0 - 100
40124
123 (7Bh)
Admin
Float
Sample pressure low alarm
Valid values are 0 - 100
40126
125 (7Dh)
Admin
Float
Sample pressure high alarm
Valid values are 0 - 100
40128
127 (7Fh)
User
16-bit
Enclosure temperature engineering unit
0=°C
1=°F
40129
128 (80h)
Read-Only
Float
Enclosure temperature low limit
Valid values are -100 - 100
40131
130 (82h)
Read-Only
Float
Enclosure temperature high limit
Valid values are -100 - 100
40133
132 (84h)
Admin
Float
Enclosure temperature low alarm
Valid values are -100 - 100
40135
134 (86h)
Admin
Float
Enclosure temperature high alarm
Valid values are -100 - 100
40142
141 (8Dh)
Admin
Float
User input 1 low alarm
Valid values are -99999.9 - 99999.9
40144
143 (8Fh)
Admin
Float
User input 1 high alarm
Valid values are -99999.9 - 99999.9
40151
150 (96h)
Admin
Float
User input 2 low alarm
Valid values are -99999.9 - 99999.9
40153
152 (98h)
Admin
Float
User input 2 high alarm
Valid values are -99999.9 - 99999.9
40160
159 (9Fh)
Admin
Float
User input 3 low alarm
Valid values are -99999.9 - 99999.9
40162
161 (A1h)
Admin
Float
User input 3 high alarm
Valid values are -99999.9 - 99999.9
40169
168 (A8h)
Admin
Float
User input 4 low alarm
Valid values are -99999.9 - 99999.9
40171
170 (AAh)
Admin
Float
User input 4 high alarm
Valid values are -99999.9 - 99999.9
Relay registers
40300
299 (12Bh)
Admin
16-bit
Measurement Hysteresis
Valid values are 0 - 65535 sec
40301
300 (12Ch)
Admin
16-bit
Time to open relay
Valid values are 0 - 2359
40302
301 (12Dh)
Admin
16-bit
Time to close relay
Valid values are 0 - 2359