Extractive installation – COSA Xentaur Xentaur XDTL-HT User Manual
Page 11

Installation
11
Extractive Installation
Enclosure
1
6
4
8
9
Calibration
Gas Inlet
Process
Gas Inlet
Sample
Gas
Exhaust
7
pipeline
Pipe Nipple
Thredolet
Ball Valve
Baseplate
Housing
Regulator
Relief Valve
Ball Valve
Outlet Connector
Pressure Gauge
Notes:
Enclosure insulated on all sides to
maintain nomially 34°C ± 1°C in a
temperature range of -20°C to 50°C.
Inlet rated to 6000 psig maximum.
Flow rate is 2 l/min nominal.
All ¼” Swaglok fittings and tubing.
Shop Air
Inlet
Shop Air
Inlet
Sample
Gas
Exhaust
5
5
2
2
10
1. Back pressure regulator
5. Adjustable regulator
9. Pressure gauge (0-30 psig)
2. Eductor
6. 3-way valve
10. Gas sampling filter
3. All ¼” stainless steel tubing
7. XTDL-HT™ laser cell
4. Flow meter
8. Enclosure heater
The instrument should be installed in a slip stream of process gas. Please make sure that:
The sample is taken from the upper surface of the main gas line to avoid problem of
contamination.
The sample is taken away from the pipe wall where flow rates may be low, and dewpoint change
may lag.
If regulators or shut-off valves are used up stream of the instrument, make sure that these do not
contain rubber or other hygroscopic materials.
Stainless steel tubing is preferred. Copper tubing is acceptable. Do not use plastic, rubber, or
Tygon tubing due to water retention in these materials.
Use small diameter pipes (1/4” or 1/8” OD).
Do not install other devices up stream of the instrument, such as other measurement systems,
which are not absolutely necessary as these are potential leak sources.