Setup Menu
Variable
Alarm State
On Error
Self Cleared
Mode
Day
Time On
Time Off
Time
Channel 1
Variable
Alarm State
On Error
Self Cleared
Variable
Alarm State
On Error
Self Cleared
Variable
Alarm State
On Error
Self Cleared
Channel 2
Channel 3
Channel 4
Relays
Channel 3 and 4 menu would not appear if the instrument is not configured with the extra relays.
Set relay X to operate on a measurement Y alarm
Go to Measurement
>
Y
>
Low Alarm
Set the alarm low threshold
Go to Measurement
>
Y
>
High Alarm
Set the alarm high threshold
Go to Relays
>
Channel X
>
Variable
Select the desired measurement
Go to Relays
>
Channel X
>
Alarm State
Select the desired alarm condition
Set relay X to operate on a system event
Go to Relays
>
Channel X
>
Variable
Select the desired event
Go to Relays
>
Channel X
>
Alarm State
Select the relay polarity
Select how relay X behaves on system fault
Go to Relays
>
Channel X
>
On Error
Turn on/off self-cleared function on relay X
Go to Relays
>
Channel X
>
Self Cleared
NOTE:
Relay 2 is reserved for controlling the enclosure heater.