Appendix B: Modbus Registers
Register
Number
Address
Access
Level
Data
Type
Description
Values
Digital input registers
40351
350 (15Eh)
Admin
16-bit
Digital input 1 mode
bit 0: Polarity (0=active high, 1=active low)
bits 1-7: input variable ID
bits 8-15: Unused
40352
351 (15Fh)
Admin
16-bit
Digital input 2 mode
bit 0: Polarity (0=active high, 1=active low)
bits 1-7: input variable ID
bits 8-15: Unused
40353
352 (160h)
Admin
16-bit
Digital input 3 mode
bit 0: Polarity (0=active high, 1=active low)
bits 1-7: input variable ID
bits 8-15: Unused
40354
353 (161h)
Admin
16-bit
Digital input 4 mode
bit 0: Polarity (0=active high, 1=active low)
bits 1-7: input variable ID
bits 8-15: Unused
Analog output registers
40401
400 (190h)
Read-Only
16-bit
Analog output 1 type
0=0-5V
1=0-10V
2=±5V
3=±10V
5=4-20mA
6=0-20mA
7=0-24mA
40402
401 (191h)
Admin
16-bit
Analog output 1 low limit DAC value
Valid values are 0 - 65535
40403
402 (192h)
Admin
16-bit
Analog output 1 high limit DAC value
Valid values are 0 - 65535
40404
403 (193h)
Admin
16-bit
Analog output 1 during calibration
operation
0=Set to low state
1=Hold output
2=Track calibration measurement