beautypg.com

COSA Xentaur Xentaur XDT User Manual

Page 25

background image

XDT User’s Manual

18

the moisture content are °C, °F, PPM, LBS and G/M

3

; the UP and DOWN buttons scroll back and

forth through these units in respective order. The °C and °F are dewpoint readings. The PPM is
parts per million by volume computed at the sensor pressure (more about pressure later). The LBS

and G/M

3

are density measurements, pounds of water per million standard cubic feet and grams

of water per standard cubic meters, respectively, both in Natural Gas, they are computed accord-
ing to data derived by IGT Research Bulletin 8, taking into account sensor pressure.
Note that the analog output is linear to the selected engineering units, therefore be mindful that
when certain units are selected, the analog output may change even though the measured dew-
point is stable. The instrument, when locked, will allow changing the engineering units only tem-
porarily and will keep the analog output linear to the original units. This feature is useful because
users can view the measurement in units other than the ones chosen for normal operation, without
disturbing the equipment connected to the output. If they forget to switch back to the original
units, the instrument will timeout and switch back - during this whole time the analog output will
operate undisturbed with the original units.
If the user desires to permanently change the units, then the instrument must be first unlocked (see
section 3.4.5-10), the units changed; it is recommended that the instrument is then locked.

The PSI legend flashes at the bottom of the display, when there is pressure correction in the com-
putation of the displayed values. A short press of the ‘pressure correct’ button toggles the unit in
and out of pressure correct mode. When there is no pressure correction applied, the PSI legend
does not appear.
Sensor Pressure is used in the context that this is the pressure inside the sample cell when per-
forming the measurement, i.e. it is the operating pressure of the sensor.
Gas Pressure is used in the context that this is the pressure at which the dewpoint is to be calcu-
lated.
Pressure Correction is used in the context that the values displayed signify the moisture content at
some pressure (we refer to this as the ‘Gas Pressure’) different from the pressure at the sensor.

Note that PPM, LBS and G/M

3

readings are by definition unaffected by pressure correction

because only the pressure at the sensor affects their value. While °C and °F are affected by pres-
sure correction by reporting what the dewpoint would be at the Gas Pressure when the dewpoint
is what is measured at the pressure at the sensor. However, this also implies that whether pressure

correction is applied or not the PPM, LBS and G/M

3

readings are affected by the setting of the

sensor pressure.
A long press of the Pressure Correct button, while in the pressure correct mode (flashing PSI leg-
end), changes the unit to the View/Set Sensor Pressure sub-state. The display has the ‘SET’ and
‘PSI’ legends on, and alternately shows SEN and the currently set value for the sensor pressure.
The up and down buttons allow the user to modify the sensor pressure, while a short press of the
pressure correct button toggles the Sensor Pressure setting between whatever value is on the dis-
play and 14.7 psi ---- this is a quick way to go back to atmospheric settings. A long press of the
pressure correct button changes the unit back to the Viewing Dewpoint Mode. Pressing the
‘Mode’ button changes the unit to the View/Set Gas Pressure sub-state. The display has the ‘SET’
and ‘PSI’ legends on, and alternately shows GAS and the currently set value for the gas pressure.
The up, down and pressure correct buttons operate in the same manner as in the Sensor Pressure
sub-state. Pressing the ‘Mode’ button changes the unit back to View/Set Sensor Pressure sub-
state, and so forth.