beautypg.com

Appendix d: hdt block diagram & connections, Hdt user’s manual page 22, Xentaur hybr id de wpoint t ransmitter – COSA Xentaur Xentaur HDT User Manual

Page 29

background image

HDT User’s Manual

page 22

Appendix D: HDT Block Diagram & Connections

Xentaur Hybr

id

De

wpoint T

ransmitter

Loop P

o

w

e

r 5V

-28V DC

, will dr

a

w

4/20mA ( pin1= +V

E

; pin2= -V

E

)

PTFE

Insulator

Stainless Steel mounting threads and housing are

electrically connected to the wider pin marked with

the ground symbol " ". They are isolated from the

current loop and sensor.

The sensor sintered filter

is electrically connected

to the loop negative pin #2.

3

1

2

Xen

tau

r

HD

T

28 V

DC

m

ax

.

+

1

3

2

Housing shield

100mA

resetable fuse

28V

3.3V

10K

47

0.001

µ

F

µ

P

0.033

µ

F

Voltage Regulator

4/20mA

Bell 202 modem

sensor

4/20mA loop

positive terminal

4/20mA loop

negative terminal

RFI filter

outer gold layer of

sensor comes in

contact with media

being measured

inner conductor

does not contact

media being

measured

stainless steel

sintered filter

connected to

loop negative