beautypg.com

COSA Xentaur Xentaur HDT User Manual

Page 15

background image

HDT User’s Manual

page 8

ppm readings are computed for a pressure of one atmo-
sphere (1 bar); and have to be corrected using a pressure
nomograph, or calculator if the sensor is measuring at
different pressures.

If readings at line pressure are necessary, it is recommended
to measure at ambient pressure and to use our dewpoint cal-
culator to obtain the dewpoint at line pressure.
Please make sure that:

The sample is taken from the upper surface of the main
gas line. This avoids problems with contamination. The
sample should be taken away from pipe line walls where
flow rates may be low, and dewpoint changes may lag.

For dewpoints dryer than -40°F, use stainless steel tub-
ing only. Copper tubing is acceptable for dewpoints wet-
ter than -40°F. Do not use plastic, rubber or tygon tubing
under any circumstances, as measurements would be
incorrect and/or response time slow due to water reten-
tion inside these materials.

Try to run pipes to the sensor upwards, so that contami-
nants tend to fall back into the main line.

Keep the length of the sample line to the sensor as short
as possible.

Use small diameter pipes (1/4” or 1/8” OD).

Use sufficient flow rates (e.g. 1 l/min with 6 feet of 1/8”
piping is adequate). The flow rate will influence the sys-
tems’ response time.

Do not install any devices upstream of the sensor, such
as other measuring systems, flow meters etc., which are
not absolutely necessary as these are potential leak
sources.

Installation of a coalescing and / or particulate filter
ahead of the sensor is desirable to prevent any liquid or
particulate contamination of the sensor.