Section 2: Sensor and Sampling Techniques

page 7

Refer to the ESS literature for more information.
If the resources to make your own sample system are avail-
able, the following two diagrams may be used as a guideline
to configure a simple system.

It is generally recommended to measure at ambient pressure
for the following reasons:

The readings will not be affected by variations in line
pressure.

The risk of exposing the sensor to liquid water is signif-
icantly reduced.

HDT

4-20mA loop cable

Exhaust

Regulator or

Needle V

alv

e

Safety shut-off Valve

Extractive Installation, Sensor

Measuring at Line Pressure

HDT

4-20mA loop cable

Exhaust

Regulator or

Needle Valve

Extractive Installation, Sensor

Measuring at Ambient Pressure

Sample Cell

Sample Cell

Main

Gas Line

Main

Gas Line

Xentaur Hybr

id

De

wpoint T

ransmitter

Loop P

o

w

e

r 5V

-28V DC

, will dr

a

w

4/20mA ( pin1= +V

E

; pin2= -V

E

)

Xentaur Hybr

id

De

wpoint T

ransmitter

Loop P

o

w

er 5V

-28V DC

, will dr

a

w

4/20mA ( pin1= +V

E

; pin2= -V

E

)