2.3.2 Extractive Installation
For extractive installations we recommend our sample sys-
tem ESS, which may be equipped with a variety of features,
such as: isolation valve, coalescing or particulate filter,
pressure regulator, calibration sample injection or extraction
port, pressure gauge, flow meter, weatherproof enclosure.
Main
Gas Line
HDT
HDT
4-20mA loop cable
Safety shut-off Valve
Bypass
Control
Valve *
Safety shut-off Valve
Bypass Installation, Sensor
Measuring at Line Pressure
In-Line Installation, Sensor
Measuring at Line Pressure
Main
Gas Line
4-20mA loop cable
NOT RECOMMENDED
Sample Cell
* maintain differential pressure to provide adequate flow through sample cell