AW Gear Meters LLC-E-BA User Manual
Page 16
AW-Lake Company 8809 Industrial Drive, Franksville, WI 53126 web: www.aw-lake.com
Tel: 262-884-9800 Fax: 262-884-9810 Email: [email protected]
REV. 1.3 LLC-E-BA Manual rev 3.3.10.DOC
16
For example:
A flowmeter has a K-factor 2053.7 imp/Gal and the display should read in Gal/Min.
KFR = 6000 / 2053.57 = 2.921741 for GPM
292174 is > 65535! 29217 > 4 digits! 2922 is best! Enter 2922 for KFR. For this
example, KFR=2.922, the decimal point is set to three decimal places in the next procedure.
Press and hold the (UP) or (DN)-key to increment or decrement the variable value. To increase
the increment or decrement rate, while holding the (UP) or (DN)-key, press and hold the (SEL)-
key for medium rate or the (ENT)-key for maximum rate. When the desired value is displayed,
press the (ENT)-key to store the programmed value and advance to the next variable.
NOTE: Always release (SEL) or (ENT)-key before releasing the (UP) or (DN)-key or the LLC
will proceed to next variable leaving the last variable with an unpredictable value.
DP for KFR - This variable selects the location of the decimal point in the KFR factor and the
display. This display will always be zeros with a decimal point.
Factory default is
00000.00
Use the (UP) or (DN)-key to move the decimal point left or right. Press the (ENT)-key to store the
location and advance to the next variable.
Eng
=g UNITS KFR - The desired engineering units for display of rate are entered using up to three
programmable characters or letters. Example: gpm
Factory default is
hz_
The first (left-most) character is programmed/edited first. To choose a letter press the (UP) or (DN)-
key until the desired character or letter appears. Press the (SEL)-key to enter the 1st character. Repeat
the procedure for the 2nd character and enter with the (SEL)-key. Repeat for the 3rd character and
store the last entry with the (ENT)-key. The next variable will display.