Van Air Systems Dew Point Meter 46-2529 2-02 User Manual
Page 5

3
3.
Connect power to terminals HOT+ and NEUT- and ground to terminal GND of TB1 as shown on
drawing.
4.
Connect Sensor Cable to terminals A and B of TB1.
5.
Recorder output signal is on TB3. Terminal 1 of TB3 is 0-5V output , Terminal 3 is
4-20mA output and Terminal 2 is signal ground.
Wiring Summary
Terminal No.TB1 Labeled Function
1
HOT+
Line Hot (or positive if 24V version board)
2
NEUT-
Line Neutral (or negative if 24V version board)
3
GND
Power and signal ground
4
NC1
Normally closed for relay #1
5
COM1
Common for relay #1
6
NO1
Normally open for relay #1
7
NC2
Normally closed for relay #2 (applies only to 6392N2)
8
COM2
Common for relay #2 (applies only to 6392N2)
9
NO2
Normally open for relay #2 (applies only to 6392N2)
10 A
Sensor
lead
11 B
Sensor
lead
12
GND
Power and signal ground
13 C
(not
used)
14 D
(not
used)
Terminal No. TB2
Function
1
Aux alarm +12VDC source
2
Aux alarm open collector output relay #1
3
Aux alarm +12VDC source (applies only to 6392N2)
4
Aux alarm open collector output relay #2 (applies only to 6392N2)
Terminal No. TB3
Function
1
Recorder output 0-5VDC
2
Power and signal ground
3
Recorder output 4-20mA
4
Aux no alarm +12VDC source
5
Aux no alarm open collector output relay #1
6
Aux no alarm +12VDC source (applies only to 6392N2)
7
Aux no alarm open collector output relay #2 (applies only to 6392N2)