Technical data, 1 general technical data – Seca 685 User Manual
Page 104

104 •
The current content of the calibration counter
flashes in the display for a few seconds.
3. Compare the content of the calibration counter out-
put with the number given on the calibration coun-
ter sticker.
Both numbers have to match for the calibration to be
valid. If the sticker and the calibration counter do not
match, the scale must be recalibrated. Please contact
your service partner or seca service. Once the scale has
been recalibrated, a new, updated calibration counter
sticker is used to identify the status of the calibration
counter. The person authorized to perform the recali-
bration secures this sticker with an additional seal. The
calibration counter sticker can be ordered from seca
service.
12.TECHNICAL DATA
12.1 General technical data
General technical data
seca 685
Dimensions of scale
• Depth
• Width
• Height
965 mm
910 mm
1100 mm
Dimensions of weighing platform
• Depth
• Width
• Height
850 mm
800 mm
57 mm
Net weight
59 kg
Temperature range
• Operation
• Storage
• Transport
+10 °C to +40 °C
-10 °C to +65 °C
-10 °C to +65 °C
Air pressure
• Operation
• Storage
• Transport
700-1060 hPa
700-1060 hPa
700-1060 hPa
Humidity
• Operation
• Storage
• Transport
30% - 80% no condensation
0% - 95% no condensation
0% - 95% no condensation
Height of digits
25 mm