Operation, 1 making the scale operational, Moving the scale – Seca 685 User Manual

Page 81: Aligning the scale

background image

Before you really get started ... • 81

English

6. OPERATION

CAUTION!

Bodily injury

Perform a function check as described in the
section entitled "Function check" on page 100
before using the device each time.

6.1

Making the scale operational

Moving the scale

1. Tilt the device until it can be moved freely on the

casters.

2. Move the device in this position to the desired

setup or storage location.

Aligning the scale

ATTENTION!

Incorrect measurement due to force shunts

If the scale and housing is in contact with some-
thing, e.g. a towel, weight will not be measured
correctly.
– Set the scale up so that only its foot screws

are in contact with the floor.

1. Place the scale on firm, level surface.

2. Undo the knurled wheels.

3. Level the device by turning the foot screws.

The air bubble in the spirit level must be located in
the exact center of the circle.

2.

3.

4.