Technical data – Seca 677 User Manual
Page 36
36
10.
Technical data
Dimensions scale
Folded up:
width
:
1089
mm
/
42.9 ins
height
:
151
mm
/
6
ins
depth
:
965
mm
/
38 ins
Set up:
width
:
951
mm
/
37.4
ins
height
:
1107
+10
mm
/
43.6
+0.4
ins
depth
:
965
mm
/
38 ins
Dimensions weighing platform
width
:
800
mm
/
31.5 ins
height
:
51
+10
mm
/
2
+0.4
ins
depth
:
965
mm
/ 38 ins
Height of figures
20
mm
Weight
approx.
34.7
kg
Temperature range
+1
0
°C
to
+40
°C
Calibration
as per Directive
2009/23/EC
approved for medical use, Class III
Medical product as per Directive
93/42/EEC
Class I
Colour
silver grey/black
Maximum load
Weighing range 1
200
kg
Weighing range 2
300
kg
Minimum load
Weighing range 1
2
kg
Weighing range 2
4
kg
Graduations
Weighing range 1
100
g
Weighing range 2
200
g
Accuracy at first calibration
Weighing range 1:
to
50
kg:
±
50
g
50
kg to 200 kg:
± 100 g
Weighing range 2:
0 to 100 kg:
± 100 g
100 kg to 300 kg:
± 200 g
Power supply
mains unit