beautypg.com

Setting up and aligning the scale, Libelle – Seca 677 User Manual

Page 25

background image

Model 677

25

GB

– Connect the mains power unit to the

scale.

– Connect the mains power unit to the

socket.

4. Setting up and aligning the scale

– Place the scale on a level surface.
– Set up the handrail as described in the

previous section.

– Undo the locknuts using the four foot

screws (1).

– Align the scale by undoing the foot

screws. The air bubble in the spirit level

must be right in the centre of the

circle (2).

– Then re-tighten the locknuts (3).

– Fold down the ramps required.
– Ensure that only the feet of the scale are

in contact with the floor. The scale may

not be in contact at any other point.

Important!

The alignment of the base of the scale must be checked and corrected if neces-
sary every time the location of the scale is changed.

1.

2.

3.

Libelle

spirit level