Correct weighing

– Press the Start key with no load on the

scale.

se(a

,

8.8.8.8.8.

and

0.0

appear consec-

utively in the display. The scale is then

automatically set to zero and ready to

use.

Switch weighing range if necessary (see

page 28).

– Stand on the scale and keep still.

Note

The maximum load on each individual

corner of the weighing platform is

200 kg.

– Read the weighing result off the digital

display.

– If a load greater than 200 kg or 300 kg

is placed on the scale,

St0P

appears in

the display.

– To switch off the scale, press the Start

key again.

Switching the weighing range

The seca 677 has two weighing ranges which can be selected. Weighing range 1 has a

reduced maximum capacity but greater display resolution, whilst weighing range 2 allows

you to use the maximum load capacity of the scale. Depending on the application in ques-
tion, you decide whether you would prefer to use the fine display resolution or the high

load range.
– Switch on the scale at the green Start

key. The scale initially operates in the

weighing range last selected.

– To change the weighing range, press

the key

.

– A triangle in the display indicates the

weighing range active in each case.

This remains active even when the scale

is turned off.