2 technical data, weighing – Seca 677 User Manual
Page 92

92 •
12.2 Technical data, weighing
Humidity
• Operation
• Storage
• Transport
30% - 80% no condensation
0% - 95% no condensation
0% - 95% no condensation
Height of digits
25 mm
Power supply
Power pack
Power consumption
• with wireless module deactivated
• with wireless module activated
approx. 20 mA
approx. 37 mA
Medical device in accordance with Directive
93/42/EEC
Class I with measuring function
EN 60601-1:
• insulated device, protection class II
• electrical medical device, type B
Wireless transmission
• Frequency band
• Transmission power
• Standards applied
2.433 GHz - 2.480 GHz
< 10 mW
EN 300 328
EN 301489-1:
EN 301489-17:
General technical data
seca 677
Technical data, weighing
Calibration in accordance with Directive
2009/23/EC
Class III
Maximum load
• Partial weighing range 1
• Partial weighing range 2
200 kg
300 kg
Minimum load
• Partial weighing range 1
• Partial weighing range 2
2.0 kg
4.0 kg
Increments
• Partial weighing range 1
• Partial weighing range 2
100 g
200 g
Tare range
300 kg
Accuracy on initial calibration
• Partial weighing range 1, to 50 kg
• Partial weighing range 1, 50 kg to 200 kg
• Partial weighing range 2, to 100 kg
• Partial weighing range 2, 100 kg to 300 kg
±50 g
±100 g
±100 g
±200 g