Seca 376 User Manual
Page 57

What to do if …? • 57
English
... if after switch-on,
measured results are
transmitted for the first
time and two acoustic
signals are heard?
• The device was unable to send measured
results to the radio receiver (seca wireless prin-
ter or PC with seca USB wireless module).
– Check that the scales are integrated in
the wireless network.
– Check that the receiver is switched on.
• Nearby RF equipment (e.g. mobile phones) are
interfering with reception.
- Make sure that RF equipment is kept at least
1 metre away from transmitters and receivers
in the seca wireless network.
NOTE:
If such interference is not eliminated, no
further acoustic warning will be given with
subsequent attempts at transmission.
... only the item "SYS" is
visible in the rf menu?
• The wireless module is disabled.
- Activate wireless module (see "Activate
wireless module (SYS)" on page 52).
... only the items "SYS"
and "lrn" are visible in
the rf menu?
• The wireless module is activated but no wireless
group set up.
- Set up wireless group (see "Set up wireless
group (Lrn)" on page 52).
... the items "APrt" and
"Time" are not visible in
the rf menu?
• No wireless printer is logged into the wireless
group.
- Use menu item "lrn" to log the wireless printer
into the wireless group (see "Set up wireless
group (Lrn)" on page 52).
…
Er:X:11
is displayed?
The scales are too high, or too great a load is
applied at one point.
- Reduce load on scales or distribute weight
more evenly
- Restart scales
…
Er:X:12
is displayed?
The scales have been switched on with too great
a load.
- Reduce load on scales
- Restart scales
…
Er:X:16
is displayed?
Oscillation of the scales has occurred, preventing
determination of the zero point.
- Do not touch the trolley or table supporting
the scales
- Restart scales
Malfunction
Cause/Remedy