Plate – Seca 376 User Manual
Page 44
44 •
Start weighing
1. Check that there is no load on the scales.
2. Press the Start key.
The display will read
se[A
and then quickly run
through all elements of the display.
The scales are ready for operation when the display
reads
0.000
.
If the device is being operated with a power supply
unit, the symbol
will appear in the display.
3. Place the baby on the scales.
4. Read off the measured result.
Tare off additional
weight (TARE)
You can use the TARE function to prevent the weighing
result from being affected by an additional weight (e.g.
a towel or pad on the weighing surface).
ATTENTION!
Incorrect measurement with force shunt
If an additional weight (e.g. large towel) is in
contact with the surface on which the scales
are standing, the weight will not be measured
correctly.
− Make sure that any additional weight is only
placed on the weighing surface of the scales.
1. Place the additional weight on the scales.
2. Hold down the arrow key (hold/tare) until "NET"
appears in the display.
3. Wait until the display stops flashing and is replaced
by
0.000
.
4. Place the baby on the scales.
5. Read off the measured result.
The additional weight has been automatically
deducted.
6. To disable the TARE function, press the arrow key
(hold/tare) until "NET" no longer appears in the
display or switch off the scales.
NOTE:
The maximum weight that can be displayed is
reduced by the weight of the objects already
placed on the scales.
hold
tare
NET
hold
tare