LumaSense Technologies INNOVA 1316A-1 User Manual
Page 79

Chapter 4
_________________________________________________________________________________________
_____________________________________________________________________________
BE6020-14
1316A-1 Multi Gas Monitor
LumaSense Technologies A/S
Instruction Manual
Page 79 of 98
After SPAN calibration revert to default mode 5. In the “Mode Select”-window set
the group “Action” to “Temporary Change” and the group “New Mode” to “5 –
6800, Single-SOL, New”, See figure 4.28. Press the “Send” button to set mode
5.
Figure 4.28 Select Mode 5 in the Mode Select window
Stop the measurement by pressing the “Stop”-button in the Main window. Finally
set the group “Zero” to "Auto" in the Main window, see figure 4.29
Figure 4.29 Set Zero calibration to Auto
This concludes the Span calibration.
4.3.6 To perform a span calibration with single component gas.
If only one gas is to be span calibrated this can be performed using a
single component gas in Nitrogen balance.
Before doing the Span calibration please make sure that Zero
calibration is set to “Manual” and the mode is set to 3 as described in