3.4.1 Pump In Flow.

Not included in the 1316A-1.

3.4.2 New NO Sensor

This fault indicate that the Full-scale output of the NO sensor is

≤80% of full scale value when sensor was installed. The NO sensor

must be replaced with a new NO sensor.

Please refer to section

4.3.8

and

5.1.8

when installing a new NO

sensor

3.4.3 New O

2

Sensor

The O

2

sensor output is detected to be below a specified limit during

Zero/O

2

Span Calibration. Replace the O

2

sensor. Please refer to

section

5.1.8

when installing a new O

2

sensor.

3.4.4 IR signal lost

This fault indicate that the sample Cell is contaminated. The sample

cell and windows must be cleaned. Might also display CO2, CO and

HC data invalid in the Data Status section of the status window see

figure 3.10

.

3.4.5 Pump Out Flow

a. Sample gas path is occluded at outlet during normal mode

operation.
Corrective action: Clear outlet occlusion. Use the pump on command

see section 5.1.3.
b. Sample gas path is occluded at the outlet during Zero/O

2

span

calibration.
Corrective action: Clear outlet occlusion. Use the pump on command

see section 5.1.3 and perform a new Zero/O

2

span calibration see

section 4.2.1.
Might also display a Zero calibration fail error in the Data status of

the status screen.
CO

2

, CO, HC, NO Zero calibration fail and O

2

span calibration fail will

be displayed.