LumaSense Technologies M360 User Manual
Page 17

M360 Manual
Installation & Operation 17
Typical Radiance temperature calibration look-up table (LUT) used for model M360
To gain access to the custom look-up table, refer to the 4 square buttons on the face of the
controller and use this procedure.
1. Press and hold 1 until ACCESS appears. If Access Level and Level 1 appears, skip to
step 3.
2. Press 4 until Level 3 appears.
3. Press 4 until 3 appears. Wait until PASS appears.
4. Press 1 until Lin16 appears.
5. Press 2 until display shows the Output # that corresponds to the calibration point shown
in the listing on the previous page.
6. Use UP/DWN buttons (3 and 4) to dial in the measured radiance temperature from a
freezing point transfer device or other radiance standard. Adjust Out# only!
7. Press 2 to view/change more correction points.
8. When finished with all calibrations and to lock the controller, press 1 until ACCESS
appears.
9. Press 3 until Level 1 appears.
Note: Calibration points do not have to be used as pre-programmed. To dial in a
different calibration point, simply change an INPUT # and the corresponding OUTPUT#
to be the new desired calibration point. Set the blackbody to the calibration
temperature using its setpoint. When exit radiance data is measured, dial in the new
OUTPUT # temperature as described above. All cal points must be in ascending order!