3. Align the connector on the rear panel of the module to the connector at the desired position of the base plate and insert the connector.

When the connectors are correctly connected, the guide pin on the rear panel of the module is inserted into the guide hole on the base
plate. Then, the module is secured to the base plate with the latch lever locking in place at the bottom section of the base plate.
Note that the main module can only be attached to the right end of the base plate.

4. For the MX100 main module, attach using a screw (M3) at the top of the module.

For the MW100 main module, attach using screws (M3) in two locations at the top of the module.

Base plate connector cover

*1

(accessory sold separately)

Input/Output module

Module attachment screw

Module connector

*2

Screw holes

Guide holes

MX100

Main module

Base plate

Base plate

Latch lever
(Pull up to remove the module from the base plate.)

Guide pins

MW100

Main module

Screw holes

Guide holes

Module connector

*2

Base plate

Base plate connector cover

*1

(accessory sold separately)

Input/Output module

Module attachment screw

Base plate

Latch lever
(Pull up to remove the module from the base plate.)

Guide pins

MX100 Main module

MW100 Main module

*1 Can be attached by hooking the top section of the cover to the top edge of the base plate and pressing down on the lower section of the

cover

When attaching the module, be sure that the module
connector is vertical to the base plate so that the
module connector pins are not bent.

*2

• Detaching the Module

Loosen the attachment screw, pull down on the latch lever on the rear panel of the module, and pull the module straight from the base
plate.

• Attachment Positions and Channel Numbers

The figure below shows how the channel numbers are identified on the PC. The MW100 channel numbers are only displayed when
viewing measured data on the MW100 Viewer software.

MX100

0

1

2

3

4

5

Slot number

001 to 010

011 to 020

021 to 030

031 to 040

041 to 050

051 to 060

Channel number in the unit

*2

Representation of
channel numbers:

Unit number

*1

Unit number

MX100: 00 to 19
MW100: 00 to 89

*2 The last one digit on a 4-channel module is 1 to 4,

the last one digit on a 6-channel module is 1 to 6,

the last one digit on a 8-channel module is 1 to 8.

Channel number in
the unit (001 to 060)

*1 Fixed to 00 when connecting using the MX100

Standard software; set between 00 and 19 when

using MXLOGGER software; and set between 00

and 89 when using the MW100 Viewer software.

Note