3 performance test, Test environment, Test instruments – Yokogawa Button Operated MV2000 User Manual
Page 18: Instrument operation, Tests, 3 performance test -3
3-3
SM MV1000-01E
Testing
1
2
3
4
5
6
7
3.3 Performance Test
Read the warning and cautions below before beginning tests.
Test Environment
Item
Conditions
Ambient temperature
23 ±5°C
Relative humidity
20 to 80% RH
Atmospheric pressure
86 kPa-106 kPa
Power supply voltage
AC power supply
(Power supply suffix code:-1) 90 to 132 VAC, 180 to 264 VAC
12 VDC power supply
(Power supply suffix code:-2) 10.0 to 28.8 VDC
Power supply frequency
Nominal frequency ±1%
Power supply waveform
Distortion factor of 5% or less
DC power supply ripple
Content ratio of 0.1% or less
Vibration
A value having negligible effects on the instrument
Position
Within ±3% of the specified position
Interference
*1
A value that does not affect the measured results
Electrical field
A value that does not affect the measured results
Magnetic field
A value that does not affect the measured results
Atmospheric pollutants
Levels of corrosive gasses, vapors, salts, and dust that do not affect
the measured results.
Other external effects
*2
A value that does not affect the measured results
*1 Interference refers to common mode noise, series mode noise, power supply noise, and
other phenomena in the signal line.
*2 Other external effects include luminance, draft, ultrasonic waves, and radiation.
Test Instruments
Instrument
Specifications
DC voltage generator
Accuracy: ±0.005% of setting + 1 µV
DMM
Accuracy: ±0.005% of rdg + 1 µV
Variable resistors
Accuracy: 0.01% + 2 mΩ or better
Insulation tester
500 VDC
Withstand voltage tester
500 to 2300 VAC, 1000 VDC
Function generator
Accuracy: ±20ppm of setting for 100 Hz
0°C standard temperature device
Accuracy: ±0.05°C
Thermocouple (Type T)
Calibrated
Instrument Operation
For operating procedures in setting mode and basic setting mode, see the following
manuals.
• MV1000 First Step Guide (IM MV1000-02E)
• MV2000 First Step Guide (IM MV2000-02E)
• MV1000/MV2000 User’s Manual (IM MV1000-01E)
Tests
The following tests are explained.
• Insulation Resistance
• Reference Junction Compensation Accuracy
• Protective Grounding
• Battery Backup
• Withstand Voltage
• Pulse Input Function (Option)