beautypg.com

3 performance test, Test environment, Test instruments – Yokogawa Button Operated MV2000 User Manual

Page 18: Instrument operation, Tests, 3 performance test -3

background image

3-3

SM MV1000-01E

Testing

1

2

3

4

5

6

7

3.3 Performance Test

Read the warning and cautions below before beginning tests.

Test Environment

Item

Conditions

Ambient temperature

23 ±5°C

Relative humidity

20 to 80% RH

Atmospheric pressure

86 kPa-106 kPa

Power supply voltage

AC power supply

(Power supply suffix code:-1) 90 to 132 VAC, 180 to 264 VAC
12 VDC power supply

(Power supply suffix code:-2) 10.0 to 28.8 VDC

Power supply frequency

Nominal frequency ±1%

Power supply waveform

Distortion factor of 5% or less

DC power supply ripple

Content ratio of 0.1% or less

Vibration

A value having negligible effects on the instrument

Position

Within ±3% of the specified position

Interference

*1

A value that does not affect the measured results

Electrical field

A value that does not affect the measured results

Magnetic field

A value that does not affect the measured results

Atmospheric pollutants

Levels of corrosive gasses, vapors, salts, and dust that do not affect

the measured results.

Other external effects

*2

A value that does not affect the measured results

*1 Interference refers to common mode noise, series mode noise, power supply noise, and

other phenomena in the signal line.

*2 Other external effects include luminance, draft, ultrasonic waves, and radiation.

Test Instruments

Instrument

Specifications

DC voltage generator

Accuracy: ±0.005% of setting + 1 µV

DMM

Accuracy: ±0.005% of rdg + 1 µV

Variable resistors

Accuracy: 0.01% + 2 mΩ or better

Insulation tester

500 VDC

Withstand voltage tester

500 to 2300 VAC, 1000 VDC

Function generator

Accuracy: ±20ppm of setting for 100 Hz

0°C standard temperature device

Accuracy: ±0.05°C

Thermocouple (Type T)

Calibrated

Instrument Operation

For operating procedures in setting mode and basic setting mode, see the following
manuals.
• MV1000 First Step Guide (IM MV1000-02E)
• MV2000 First Step Guide (IM MV2000-02E)
• MV1000/MV2000 User’s Manual (IM MV1000-01E)

Tests

The following tests are explained.
• Insulation Resistance

• Reference Junction Compensation Accuracy

• Protective Grounding

• Battery Backup

• Withstand Voltage

• Pulse Input Function (Option)

• Measurement Accuracy

This manual is related to the following products: