3.2 Self Diagnostic Test
The MV is provided with complete self diagnostic functions to enhance reliability in
measurement and serviceability.
When you turn ON the power, the MV will automatically execute the following types
of diagnoses alternately and display the results. After these tests are completed, the
recorder is ready for use.
1.
Main program (Flash ROM) test
2.
Main RAM write/read test
3.
A/D and A/D EEPROM test
4.
Memory acquisition test (write test from Flash)
5.
Ethernet module test
The table below shows the results of the self diagnostic tests when a problem is
detected.
Display upon Error
Likely Cause
Refer to Chapter 6
Part Name
MV1000
MV2000
Page Item Page Item
901
902
ROM failure.
RAM failure.
Connection between MAIN BOARD
ASSEMBLY and SUB BOARD
ASSEMBLY
–
–
–
–
–
Replace the SUB BOARD ASSEMBLY
or MAIN BOARD ASSEMBLY
Sub PCB Assembly
6-3
5
6-12 13
Main PBA
6-4
12 6-12 11
910
921
A/D memory failure for
all input channels.
A/D calibration value
error.
Connection between AD SCANNER
BOARD ASSEMBLY and MAIN BOARD
ASSEMBLY
–
–
–
–
–
Replace the AD SCANNER BOARD
ASSEMBLY or
MAIN BOARD ASSEMBLY
AD-STD/ISO
6-3 7, 8
6-12
24,
25,
26,
27
Main PBA
6-4
12 6-12 11
930 Memory acquisition
failure.
Connection between MAIN BOARD
ASSEMBLY and SUB BOARD
ASSEMBLY
–
–
–
–
–
Replace the SUB BOARD ASSEMBLY
or MAIN BOARD ASSEMBLY
Sub PCB Assembly
6-3
5
6-12 13
Main PBA
6-4
12 6-12 11
Replace the internal CF card
–
6-2 –
6-10 –
940 The Ethernet module
is down.
Check communication settings
–
–
–
–
–
Connection between POW TERM &
COMM BOARD ASSEMBLY and MAIN
BOARD ASSEMBLY
–
–
–
–
–
Check the communication cables and
LINK LED illumination
–
–
–
–
–
Replace the MAIN BOARD ASSEMBLY
or POW TERM & COMM BOARD
ASSEMBLY
Main PBA
6-4
12 6-12 11
I-PWR Terminal
24V-Power Terminal
Power Terminal
6-4
8
6-14
3